ZX1030E AC/DC Single Phase Standard Power Source
This device is used for testing and calibrating all kinds of ammeter, voltmeter, ohmmeter under 0.2 level. It can also be used as testing power source, matching with high level standard meters to calibrate ammeter, voltmeter, ohmmeter at 0.2 or higher level. There are two regulation modes: analog regulation (potentiometer) and digital regulation (keyboard).
1. 5½ LED digitally display output by actual value or by percentage
2. AC/DC current output range: 0~100 uA ~500uA ~2mA~5mA~20mA ~50mA ~200mA ~500mA ~2A~5A~20A~30A(optional)~50A(optional)
3. AC/DC voltage output range: 0~200mV~1V~2V~5V~10V~20V~50V~100V~200V~500V~1000V
4. Three AC frequency for option: 50Hz、60 Hz、400 Hz
5. Auto protection or manual resetting function for over load
6. Over load rang is 10% for each level of voltage, current
7. Clip-on transformer testing: matching with our product of standard coil, measuring range AC 0~750A ( above 1000A for part of meters) DC 0~1000A, error: ±0.3%
1. Stability: DC < 0.01% of full range / 2 min, AC < 0.03% of full range / 2 min
2. AC distortion: ﹤0.5%
3. DC ripple coefficient: ﹤0.1%
4. Output frequency and accuracy: 50 Hz、60 Hz、400 Hz: ﹤1%
5. Voltage Current Output range and accuracy: (23℃±2℃ > 10% range)
6. Power supply consumption: AC DC 220V±10%, 50 Hz±1Hz, ﹤180VA
7. Working environment: temperature 5℃~35℃ relative humidity ≤80%
8. Working time: continuous
9. Size and weight: 135×480×420mm 21KG

-
Pre-sale service commitments
-
Sale service commitments
-
Sale service commitments
- ZX1030: A Reliable Single-Phase Power Calibrator for Everyday Meter Testing2026-08-10
- Precision in Power Testing: The ZX3030X Harmonic Standard Source2025-10-11
- Precision Power Calibration: The ZX3030H Three-Phase Standard Source2025-10-10
- Advanced Testing for Substation Automation: ZX3030F vs. Traditional Methods2025-10-09



Manual
Video