ZX3030A Three Phase AC Program control Standard Power Source
The instrument is featured with meter-power combination structure, full digital closed loop technology, embedded system, large TFT color LCD, CPLD chip, and advanced A/D converters, little weight and small volume. It is widely used in electric energy metering industry, electrical laboratory and other relevant departments
1. Design with touch panel and mask button, massive information to see and simple to use
2. built-in standard energy meter for digital closed loop feedback, U, I, Φ, F, W, VAR, VA, PF all for reference
3. Perfect over circuit, over voltage. short circuit, open circuit, over heat, over load protection
4. Three phase rated current rated voltage can be preset by users
5. Multiple choose of voltage, current, phase testing point
6. Capable of measuring single phase, three phase, active power energy meters and reactive power energy meters
7. Large memory for data storage, and searching data is very convenient
8. With R232 communication interface for PC control and software upgrade
1. AC Phase voltage output
① Regulation fineness: 0.01% RG
② Resolution: 6 effetive digits
③ Accuracy: better than 0.05% RG
④ Stability: better than 0.01%RG/ 1 min
⑤ Distortion degree: <0.1% (non capacitive load)
⑥ Output power: rated 25 VA/phase
⑦ Full load regulation rate: ≤0.01% RG
⑧ Range: 10V - 500V
⑨ Four gears: 57.7 V, 100 V, 220 V, 380 V, switch automatically
⑩ Long-term stability: ± 60 PPM/year
2. AC current output
① Adjust fineness: 0.01% RG
② Accuracy: better than 0.05% RG
③ Resolution: 6 effetive digits
④ Stability: <0.02% RG / 1 min
⑤ Distortion degree:< 0.1%
⑥ Output power: rated output 20 VA/Phase
⑦ Full load regulation rate: 0.01% RG
⑧ Range: 0.5mA~24A
⑨ Gears: 0.2A, 1A, 5A, 20A, switch automatically
3. Power output
① Accuracy: better than 0.05% RG
② Stability: better than 0.01% RG / 1 min
③ Resolution: 6 effective digits ( active power, reactive power, apparent power)
4. Power factor
① Adjusting range:-1 ~ 0 ~ + 1
② Resolution: 0.00001;
③ Accuracy: 0.0005
5. Phase angle
① regulation range: 0 ~ 359.99°
② Resolution: 0.001°
③ Accuracy: 0.02°
6. Frequency
① Adjusting range: 40 Hz ~ 70 Hz
② Resolution: 0.001 Hz
③ Accuracy: 0.002 Hz
7. Harmonic output: 2~51 orders
-
Pre-sale service commitments
-
Sale service commitments
-
Sale service commitments